Contact an Account ManagerAnalysis of organochlorine pesticides in accordance with EPA 8081 on mid-polar ultra low bleed column
Summary
Type: GC
Application #: 213630
Catalog#: 726162.30 - GC Column, OPTIMA-17 MS, 0.25mm, 0.25um, 30m
Detection:
Conditions: Injection volume: 1 µl, 0.5 min splitless, 25 ml/min
Injection temperature: 280 °C
Carrier gas: Helium, 0.12 bar (1.5 ml/min)
Temperature: 100 °C for 0.5 min --> 210 °C, 40 °C/min --> 250 °C, 6 °C/min -->
330 °C, 15 °C/min, for 5 min
Authors: GC department
Source: Macherey-Nagel; Germany 2010
Details
Legend: