106790 Rapid and sensitive method for HPLC analysis of pterins in biological fluids
Home
sorbtech.com
Back
Contact an Account Manager
Rapid and sensitive method for HPLC analysis of pterins in biological fluids
Summary
Type:
HPLC
Application #:
106790
Detection:
fluorescence, excitation 350nm, emission 410nm
Conditions:
eluent: water - methanol (97:3, v/v)
flow rate 0.7ml/min
temperature 35°C
Authors:
Antonozzi I
Source:
J Chromatogr 459 (1988) 319 - 324
Details
Legend: Fig. 1. Chromatograms for a) standard mixture of biopterin (B), neopterin (N), pterin (Pt) and 6-methylpterin (6-mPt), b) urine sample treated with acidic oxidation and c) the same urine sample trea