103710 Determination of teicoplanin in human plasma and urine by affinity and reversed phase HPLC
Home
sorbtech.com
Back
Contact an Account Manager
Determination of teicoplanin in human plasma and urine by affinity and reversed phase HPLC
Summary
Type:
HPLC
Application #:
103710
Catalog#:
720141.46 - MN HPLC Column, Nucleosil C18, 100A, 5um, 4.6x100mm
Detection:
UV, 240nm
Conditions:
eluent A: 0.01M NaH2PO4 pH 4.9
eluent B: acetonitrile - eluent A (50:50, v/v)
flow rate 1.3ml/min
Authors:
Riva E
Source:
J Chromatogr 421 (1987) 99 - 110
Details
Legend: Fig. 2. Typical chromatogram obtained from a standard solution of teicoplanin (100µg/ml pure substance), and the gradient profile (% B) used for the elution of components, column cleanup and reequil